×

Cardiac pacemaker with hysteresis behavior

  • US 5,284,491 A
  • Filed: 02/27/1992
  • Issued: 02/08/1994
  • Est. Priority Date: 02/27/1992
  • Status: Expired
First Claim
Patent Images

1. In a pacemaker having a programmable lower rate and upper rate, a programmable lower rate hysteresis (LRH) corresponding to a lower rate hysteresis interval (LRHI), and a programmable intermediate pacing rate (IR) the improvement comprising:

  • A. means for measuring rate of change MAVG of successive cardiac intrinsic escape intervals;

    B. means for comparing MAVG to a first predefined limit SL1 ;

    C. means for comparing the last intrinsic escape interval to the lower rate hysteresis interval (LRHI); and

    D. means for stimulating a heart at the lower rate hysteresis (LRH) and for gradually incrementing a pacing rate until the pacing ratio reaches the intermediate pacing rate (IR) if the last intrinsic escape interval is longer than the lower rate hysteresis interval (LRHI) and if said MAVG is greater than SL1.

View all claims
  • 1 Assignment
Timeline View
Assignment View
    ×
    ×