Method of determining calibration curve and analysis method and apparatus using the same
First Claim
1. An apparatus for quantitative analysis of an analyte in a sample by referring to a calibration curve prepared by using the interrelation between the known concentrations of the analyte in standard samples and analytical measured values of the standard samples, said apparatus comprising:
- 1) input means for inputting the known concentrations (pi) of the analytes contained in plural standard samples and their analytical measured values (qi);
2) a processor for receiving the data input from said input means and processing the operations mentioned below to determine a regression function of the calibration curve on the basis of the input (pi) and (qi), said operations including steps of;
(a) expressing the part of calibration curve of an intermediate concentration region (p1 -p2), which is defined as the region of linear portion in a semi-logarithmic graph in which respective concentrations (pi) of the standard sample and the logarithmic values (log qi) of the corresponding analytical measured values (qi) are plotted on a rectangular coordinate system, said linear portion having the most proximate concentrations p1 and p2 at both ends, as
space="preserve" listing-type="equation">Y=exp (b·
X+d) which means
space="preserve" listing-type="equation">X=(ln Y-d)/b (1) wherein Y is the analytical measured value (qi),X is the concentration of the analyte in the standard sample (pi), andb and d are coefficients;
(b) expressing the part of calibration curve of the low concentration region (from the minimum concentration p0 in the standard sample to concentration p1) as
space="preserve" listing-type="equation">X=e·
Y.sup.2 +f·
Y+g (2) wherein e, f and g are coefficients;
(c) expressing the part of calibration curve of the high concentration region (from concentration p2 to the maximum concentration p3 in the standard sample) as
space="preserve" listing-type="equation">X=l·
Y.sup.2 +m·
Y+n (3) wherein l, m and n are coefficients;
(d) setting boundary conditions where the differentiated value (dX/dY) of Equation (1) at the coordinate (p1, q1) is equal to the differentiated value (dX/dY) of Equation (2) at the coordinate (p1, q1) and the differentiated value (dX/dY) of Equation (2) at the coordinate (p2, q2) is equal to the differentiated value (dX/dY) of Equation (3) at the coordinate (p2, q2); and
(e) calculating respective coefficients in Equations (1), (2) and (3) and finding therefrom the continuous regression function of the calibration curve for the whole concentration region covering from p0 to p3 ;
3) a calibration curve generator for generating the calibration curve to be used for quantitative analysis of the analyte contained in a test sample, the calibration curve being prepared from the regression function determined by said processor;
4) a measurement means for measuring an analytical value of the sample containing an unknown concentration of the analyte to obtain an analytical measured value; and
5) an output means for calculating and outputting the concentration of the analyte in the sample by comparing the analytical measured value of the sample and the calibration curve.
0 Assignments
0 Petitions
Accused Products
Abstract
One sigmoid calibration curve is split into three parts of low concentration region represented by a high degree function, intermediate concentration region represented by an exponential function and high concentration region represented by a high degree function according to the present invention. The boundary condition of the adjacent two functions is set so that the two functions have an equal slope at the boundary point; thereby, regression functions of the calibration curves in respective regions are found. The number of standard samples for finding a calibration curve can be reduced while the calibration curve found is of high accuracy.
17 Citations
8 Claims
-
1. An apparatus for quantitative analysis of an analyte in a sample by referring to a calibration curve prepared by using the interrelation between the known concentrations of the analyte in standard samples and analytical measured values of the standard samples, said apparatus comprising:
-
1) input means for inputting the known concentrations (pi) of the analytes contained in plural standard samples and their analytical measured values (qi); 2) a processor for receiving the data input from said input means and processing the operations mentioned below to determine a regression function of the calibration curve on the basis of the input (pi) and (qi), said operations including steps of; (a) expressing the part of calibration curve of an intermediate concentration region (p1 -p2), which is defined as the region of linear portion in a semi-logarithmic graph in which respective concentrations (pi) of the standard sample and the logarithmic values (log qi) of the corresponding analytical measured values (qi) are plotted on a rectangular coordinate system, said linear portion having the most proximate concentrations p1 and p2 at both ends, as
space="preserve" listing-type="equation">Y=exp (b·
X+d)which means
space="preserve" listing-type="equation">X=(ln Y-d)/b (1)wherein Y is the analytical measured value (qi), X is the concentration of the analyte in the standard sample (pi), and b and d are coefficients; (b) expressing the part of calibration curve of the low concentration region (from the minimum concentration p0 in the standard sample to concentration p1) as
space="preserve" listing-type="equation">X=e·
Y.sup.2 +f·
Y+g (2)wherein e, f and g are coefficients; (c) expressing the part of calibration curve of the high concentration region (from concentration p2 to the maximum concentration p3 in the standard sample) as
space="preserve" listing-type="equation">X=l·
Y.sup.2 +m·
Y+n (3)wherein l, m and n are coefficients; (d) setting boundary conditions where the differentiated value (dX/dY) of Equation (1) at the coordinate (p1, q1) is equal to the differentiated value (dX/dY) of Equation (2) at the coordinate (p1, q1) and the differentiated value (dX/dY) of Equation (2) at the coordinate (p2, q2) is equal to the differentiated value (dX/dY) of Equation (3) at the coordinate (p2, q2); and (e) calculating respective coefficients in Equations (1), (2) and (3) and finding therefrom the continuous regression function of the calibration curve for the whole concentration region covering from p0 to p3 ; 3) a calibration curve generator for generating the calibration curve to be used for quantitative analysis of the analyte contained in a test sample, the calibration curve being prepared from the regression function determined by said processor; 4) a measurement means for measuring an analytical value of the sample containing an unknown concentration of the analyte to obtain an analytical measured value; and 5) an output means for calculating and outputting the concentration of the analyte in the sample by comparing the analytical measured value of the sample and the calibration curve. - View Dependent Claims (2, 3, 4)
-
-
5. An apparatus for quantitative analysis of an analyte in a sample by referring to a calibration curve prepared by using the interrelation between the known concentrations of the analyte in standard samples and analytical measured values of the standard samples, said apparatus comprising:
-
1) input means for inputting the known concentrations (pi) of the analytes contained in plural standard samples and their analytical measured values (qi); 2) a processor for receiving the data input from said input means and processing operations to determine a regression function of the calibration curve on the basis of the input (pi) and (qi), said operations comprising; (a) expressing the part of calibration curve of an intermediate concentration region (p1 -p2) as an exponential function, said intermediate concentration region being defined as the linear portion of a semi-logarithmic plot of respective concentrations (pi) of the standard sample and the logarithmic values (log qi) of the corresponding analytical measured values (qi) on a rectangular coordinate system; (b) expressing the part of calibration curve of a low concentration region as a multi-degree function, said low concentration region being defined as that portion of the semi-logarithmic plot from the minimum concentration (p0) to the lower concentration end (p1) of the intermediate region; (c) expressing the part of calibration curve of the high concentration region as a multi-degree function, said high concentration region being defined as that portion of the semi-logarithmic plot from the higher concentration end (p2) of the intermediate concentration region to the maximum concentration (p3); (d) setting boundary conditions that at the boundary points between the multi-degree functions and the exponential function, the values of the first derivative of the multi-degree function and the first derivative of the exponential function are identical; and (e) calculating the continuous regression function by determining the identity of the functions for the intermediate, low and high concentration regions which correspond to requirements (a) through (d); 3) a calibration curve generator for generating the calibration curve to be used for quantitative analysis of the analyte contained in a test sample, the calibration curve being prepared from the regression function determined by said processor; 4) a measurement means for measuring an analytical value of the sample containing an unknown concentration of the analyte to obtain an analytical measured value; and 5) an output means for calculating and outputting the concentration of the analyte in the sample by comparing the analytical measured value of the sample and the calibration curve. - View Dependent Claims (6, 7, 8)
-
Specification