×

Device and method for using oscillatory pressure ratio as an indicator for lung opening during high frequency oscillatory ventilation

  • US 6,390,092 B1
  • Filed: 08/07/2000
  • Issued: 05/21/2002
  • Est. Priority Date: 08/07/2000
  • Status: Expired due to Term
First Claim
Patent Images

1. A method of determining the optimum mean airway pressure for a patient on high frequency oscillatory ventilation comprising the steps of:

  • (a) measuring the peak-to-peak oscillatory pressures in the proximal and distal ends of an endotracheal tube positioned within the patient;

    (b) calculating the oscillatory pressure ratio from the peak-to-peak oscillatory pressures in the proximal and distal ends of the endotracheal tube;

    (c) altering the mean airway pressure to the patient;

    (d) recalculating the oscillatory pressure ratio at the altered mean airway pressure; and

    (e) repeating steps (c) and (d) until the oscillatory pressure ratio is at or near its minimum value.

View all claims
  • 4 Assignments
Timeline View
Assignment View
    ×
    ×